DX-D 800
Fluoroscopy parameters:
•
From 40 to 120 kV, by steps of 1 kV,
accuracy ± (3 %+ 1 kV)
•
From 0.5 to 5 mA (low dose) and up to 7 mA
(
higher dose)
Operating modes:
• 999
customizable anatomical protocols
• 3
points (kV, mA, s)
• 2
points (kV, mAs)
• 1
point (kV, AEC)
Dimensions of active areas: 2 x (90 x 40) mm+
1
x (100 x 40) mm.
Sensitivity difference between sensor fields: < 10 %.
X-RAY TUBE
Tube speci cation:
•
Nominal focal spot values: 0.6 and 1.2 mm
•
Anode rotation speed: 9,000 rpm
•
Nominal anode input power: 43 and 100 kilowatts
•
Anode diameter: 102 mm
•
Anode material: Rhenium - Tungsten - Molybdenum
•
Anode angle: 13°
•
X-ray coverage at 1 m: 43 x 43 cm
•
Inherent filtration: 0.7 mm
•
Radiation protection: complies with IEC 60-601-3
standard
•
Anode heat storage capacity: 600 kHU
•
Anode heat dissipation rate: 1,000 W (81.1 kHU/min;
60
kJ/min)
•
Housing: C 100 XT
•
House heat storage capacity: 2,025 kHU (1,000 kJ)
•
Housing heat dissipation rate: 1,000 W (81 kHU/min;
60
kJ/min) with fan
•
Cooling: Air
•
High Voltage cable length: 12 m
DOSE AREA PRODUCT MEASURING SYSTEM
•
Active area: 147 x 147 mm
•
Stabilization time: 6 minutes
•
Measuring range: from 10 to 99,999,999 mGy.cm²
•
Digital resolution: 1 mGy.cm²
•
Sensitivity uniformity: < 6 %
•
Accuracy: ± 25 %
•
Filtration: Eq Al (70 kV) 0.2 mm
•
Transparency: > 70 %
REMOTE CONTROLLED TABLE
Speci cations:
•
Tabletop dimensions: 2,250 x 810 mm (600 mm, net width)
•
Ground to tabletop distance: 640 to 930 mm
•
Radiation attenuation: 0.4 mm eq. Al
•
Lateral movement: motorized at 4 cm/s
•
Clearance: +180 mm/-180 mm
•
Longitudinal movement: motorized at 4.35 cm/s
•
Clearance: +1,200 mm/-600 mm or
+600 mm/-1,200 mm
Software provides anti-collision safety.
Tilting:
•
Angulation: +90°/-25° ± 0.5° (Trendelenburg)
•
Tilting time: 15 seconds from 0° to 90° (± 2°)
Software provides anti-collision safety.
Micro switches provide mechanical safety.
Maximum patient’s weight:
•
Maximum allowed patient’s weight on the table: 230 kg
(500
lbs) without any movement restriction
•
Maximum allowed patient’s weight on the foot rest:
150
kg (330 lbs)
Tube support:
•
Motorized collimator fitted with four pairs of lead
shutters, and 100 W halogen bulb and laser; manual
rotation of ± 30° possible
•
CANbus
•
Automatic, semi-automatic and manual modes
•
Automatic mode: shutters are opened based on detector
format, anatomical region (APR), auto-positioning or
selected fluoro field
•
Manually: the collimation can be adjusted more precisely
to the area to be X-rayed from the main console using the
proportional joystick, with the remote hand switch or
from the collimator
•
For dose reduction, automatic collimation and filters
are predetermined in each protocol; video camera for
positioning the patient without irradiation. The video
images are displayed on the touch screen of the main table
console
•
Motorized source-image distance: 6 cm/s, variable from
110
to 180 cm