DX-D 800
5
SYSTEM ACCESSORIES AND OPTIONS
Optional:
•
Tomography
•
One direction curvilinear tomography
•
Angles: 40°, 20°, 10°
• 2
speeds per angle:
0.5
s max for a 10° angle
1
s max for a 20° angle
2
s max for a 40° angle
•
Layer height: from 0 to 300 mm
•
Layer height value displayed on the control desk
•
Automatic layer height increase or decrease
•
Software for anti-collision safety
•
Micro switches for mechanical safety
Optional:
Motorized and removable compression device.
Patient safety system inhibits tabletop and angulation
movements when compression is in use. The manual
compression device is composed of two winches and
one belt.
Optional:
•
Automatic stitching (image pasting) software to
install on console
•
Lateral holder for portable detector
•
Shoulder rest, gynecological stirrups, etc.
•
One composite footrest, one adjustable stool, two
patient handles, two winches & one belt
EQUIPMENT ENVIRONMENT
Dimensions (l x w x h) and weight
•
Table: 2,260 x 1,590 x 2,925 mm (88.97 x 62.60 x
115.15
inch) maximum, in horizontal position, higher
height, SID at 180 cm and column at 0°, 980 kg
(2,160.53
lbs)
•
Table cabinet integrated into the table
•
Table main console: 600 x 300 x 150 mm (23.62 x
11.81
x 5.91 inch), 6.5 kg (14.33 lbs)
•
X-ray tube: 524 x Ø 190 mm (20.63 x Ø 7.48 inch),
29
kg (63.93 lbs)
•
Generator cabinet: 592 x 360 x 690 mm (23.31 x 14.17 x
27.16
inch), 95 kg (209.44 lbs)
•
Generator control interface integrated into table console
•
Flat panel detector: 493 x 503 x 26 mm (19.41 x 19.80 x
1.02
inch), 6 kg (13.23 lbs)
•
Acquisition console/computer: 360 x 310 x 100 mm
(14.17
x 12.20 x 3.94 inch), 9.8 kg (21.61 lbs)
•
Acquisition console/color LCD monitor: 474 x 470 x
220
mm (18.66 x 18.50 x 8.66 inch), 11.5 kg (25.35 lbs)
•
LCD monitor in room on trolley: 432 x 251 x 467 mm
(17.01
x 9.88 x 18.39 inch), 10.9 kg (24.03 lbs)
Operating environmental conditions
:
•
Temperature: from 10° C to 35° C
(
the temperature must change progressively)
•
Relative humidity: from 30 to 75 % (no condensing)
•
Atmospheric pressure: from 700 to 1,060 HPa
Power supply:
•
Generator: 400 V ac ± 10 %, three phases, 50/60 Hz
•
Table: 400 V ac ± 15 V, three phases + ground +
impedance neutral, 50/60 Hz
•
Digital system by FPD: 230 V ac ± 10 %, single phase,
50/60
Hz
Protection:
•
Generator: 63 A circuit breaker, D power line, with
30
mA differential sensitivity
•
Table: 12 A circuit breaker, D power line, with 30 mA
differential sensitivity
•
Digital system by FPD: 2 x 16 A circuit breaker, D power
line, with 30 mA differential sensitivity