4
DX-D 400
COLLIMATORS
Inherent filtration
2
mm aluminum
equivalent
Full field light localizer
200
lx
Additional filtration
1
mm Al + 0.1 mm Cu
1
mm Al + 0.2 mm Cu
2
mm Al or 1 mm Al
Rotation
up to maximum
± 90°
WALL STAND
Dimensions
Height
2235
mm (87.99 inch)
Width
620
mm (24.41 inch)
Length
345
mm (13.58 inch)
Height from horizontal axis of film
Minimum height
400
mm (15.75 inch)
Maximum height
1900
mm (74.8 inch)
Bucky vertical motion
1500
mm (59.06 inch)
Table surface-film
plane distance
44
mm (1.73 inch)
Radiation absorption
< 0.8 mm Al
equivalent
Automatic exposure control
3-
field solid state
sensors
Scatter radiation grid
150/180
cm option
(59.06/70.87
inch)
SYSTEM ACCESSORIES
Compression band
Hand grips for wall stand
Hands support for wall stand
Lateral cassette holder 24 x 30
Lateral cassette holder 35 x 43
Vacudap 2004 (external dose area meter)
INSTALLATION DATA
Line voltage
3-
phase,
380/400/415/440
and
480
v 50/60 Hz
(
additional transformer required for 80 kW generator
below 480 v)
Automatic line compensation
± 10 %
Power consumption
105
kVA (Gen 64 kW);
120
kVA (Gen 80 kW)
Radiographic Table
Single phase 110 -- 230
V~, 50/60 Hz
Ceiling height
2.60 - 2.80
m (102.36 -
110.24
inch)
(
for normal use)
Ambient conditions (operation)
Temperature range
+15° C to +30° C
Storage/Transport temperature -20° C to +70° C
Relative humidity
(
no condensing)
30 %
to 75 %
Atmospheric pressure
700
hPA to 1060 hPa
PRODUCT WEIGHTS
Patient table: non elevating:
190
kg (418.88 lbs)
Patient table: elevating:
280
kg (617.29 lbs)
Floor mounted tubestand:
345
kg (760.59 lbs)
Wall stand:
145
kg (319.67 lbs)