DX-D 400
5
X-RAY TUBES
Housing
Focal spot
Target angle
Heat capacity (kHU)
Anode speed
Toshiba E7884X
0.6 – 1.2
12
°
300
Low
Toshiba E7252X
0.6 – 1.2
12
°
300
High/Low
Toshiba E7254FX
0.6 – 1.2
12
°
400
High/Low
Toshiba E7869X
0.6 – 1.2
12
°
600
High/Low
GENERATORS
Generator model
SHF 335/345
SHF 535/545
SHF 635/645
SHF 835/845
(
CR only)
Input power
3
Phase 380 (x35)
or 480 (x45)
Max. power (kW)
32
50
64
80
Max. mA
400
640
640
800
Power output
640
mA @ 78 kVp
640
mA @ 80 kVp
800
mA @ 80 kVp
@ 0.1
s
500
mA @ 100 kVp
640
mA @ 100 kVp
800
mA @ 100 kVp
400
mA @ 125 kVp
500
mA @ 128 kVp
640
mA @ 125 kVp
320
mA @ 150 kVp
400
mA @ 150 kVp
500
mA @ 150 kVp
Compatible X-ray
All
All
E7252X, E7254FX,
E7254FX, E7869X
Tubes
E7869X
Range of radiographic parameters
mA
From 10 mA to 800 mA through the following mA stations: 10, 12.5, 16, 20, 25, 32, 40, 50, 64,
80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800.(
Depending on the Generator model)
mAs
Product of mA x time values from 0.1 mAs to 500 mAs
ms
From 1 to 10000 milliseconds through the following time stations: 1, 2, 3, 4, 5, 6, 8, 10, 12, 16,
20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800, 1000, 1250, 1600,
2000, 2500, 3200, 4000, 5000. 6400, 8000
and 10000.
AEC (Automatic
mAs: 0.1 mAs to 500 mAs exposure time: Nominal shortest irradiation time = 1 ms
Exposure Control)